iHealth Align User Manual
Page 12
Cleaning And Disinfecting the Meter and Lancing Device
The cleaning and disinfection is absolutely necessary for the test procedure, because cleaning
can ensure the meter works well (for example, display will be clear to see after cleaning); and
disinfection will reduce the chance of spreading infection to others.
The meter and lancing device should be cleaned and disinfected following each use. We suggest
that you use CaviWipes™ (Metrex® Research Corporation, EPA Reg. No. 46781-8, EPA Est. No.
Signs of Potential Physical And Performance Deterioration
If you encounter any of the following circumstances, stop using the meter and contact the place
of purchase for assistance, or call the iHealth toll free customer service number at
1-855-816-7705:
1. The device does not work; for example, the mobile device does not begin testing when the
meter is connected with the mobile device and when a test strip is inserted into the meter.
2. Discoloration of the meter casing or lancing device; for example, it is difficult to read the
labeling information.
3. Corrosion, crazing (fine cracks), embrittlement, and/or cracking of the meter casing or lancing
device.
10
56952-WI-001). CaviWipes, with isopropanol and diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl
ammonium chloride as the active ingredients, have been shown to be safe for use with the meter
and lancing device.
You can purchase this product from the suppliers listed below:
(1) Visit the website www.metrex.com or contact Metrex at 800-841-1428 for product or technical
information.
(2) Visit the website http://www.endochoice.com/Equipment?search=wipe.
The meter and lancing device are validated to withstand 11,000 individual tests and 11,000
cleanings over their 5-year life spans.
Steps for cleaning and disinfecting the meter and lancing device
1. After a test, clean and wash your hands.
2. Use one CaviWipe to carefully clean the meter, front and back.
3. Disinfect the meter with another wipe and let it air dry.
4. Use the same method with the CaviWipes to clean and disinfect the lancing device.
5. Take off and discard the sanitary cover for the mobile device after each test.
WARNING
①
Each disinfection step requires a pre-cleaning step. Wash hands thoroughly with soap and
water after handling the meter, lancing device, or test strips.
②
Only the surface of the meter can be cleaned and disinfected with the disinfecting towelette.
Do not insert the CaviWipe into the test port.