beautypg.com

iHealth Align User Manual

Page 12

background image

Cleaning And Disinfecting the Meter and Lancing Device

The cleaning and disinfection is absolutely necessary for the test procedure, because cleaning

can ensure the meter works well (for example, display will be clear to see after cleaning); and

disinfection will reduce the chance of spreading infection to others.

The meter and lancing device should be cleaned and disinfected following each use. We suggest

that you use CaviWipes™ (Metrex® Research Corporation, EPA Reg. No. 46781-8, EPA Est. No.

Signs of Potential Physical And Performance Deterioration

If you encounter any of the following circumstances, stop using the meter and contact the place

of purchase for assistance, or call the iHealth toll free customer service number at

1-855-816-7705:

1. The device does not work; for example, the mobile device does not begin testing when the

meter is connected with the mobile device and when a test strip is inserted into the meter.

2. Discoloration of the meter casing or lancing device; for example, it is difficult to read the

labeling information.

3. Corrosion, crazing (fine cracks), embrittlement, and/or cracking of the meter casing or lancing

device.

10

56952-WI-001). CaviWipes, with isopropanol and diisobutyl-phenoxy-ethoxyethyl dimethyl benzyl

ammonium chloride as the active ingredients, have been shown to be safe for use with the meter

and lancing device.

You can purchase this product from the suppliers listed below:

(1) Visit the website www.metrex.com or contact Metrex at 800-841-1428 for product or technical

information.

(2) Visit the website http://www.endochoice.com/Equipment?search=wipe.

The meter and lancing device are validated to withstand 11,000 individual tests and 11,000

cleanings over their 5-year life spans.

Steps for cleaning and disinfecting the meter and lancing device

1. After a test, clean and wash your hands.

2. Use one CaviWipe to carefully clean the meter, front and back.

3. Disinfect the meter with another wipe and let it air dry.

4. Use the same method with the CaviWipes to clean and disinfect the lancing device.

5. Take off and discard the sanitary cover for the mobile device after each test.

WARNING

Each disinfection step requires a pre-cleaning step. Wash hands thoroughly with soap and

water after handling the meter, lancing device, or test strips.

Only the surface of the meter can be cleaned and disinfected with the disinfecting towelette.

Do not insert the CaviWipe into the test port.