Inova Labs XYC100B-SPT User Manual
Page 31

29
The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-
SPT/XYC100B-PRO is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user
of the Model XYC100B-SPT/XYC100B-PRO can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Model XYC100B-SPT/XYC100B-
PRO as recommended below, according to the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance ‘d’ in meters can be estimated using the equation applicable
to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
Recommended Separation Distances Between Portable and Mobile
RF Communications Equipment and The LifeChoice
®
Activox Portable
Oxygen Concentrator Model XYC100B-SPT/XYC100B-PRO
The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-SPT/
XYC100B-PRO is intended for use in the electromagnetic environment specified
below. The customer or the user of the Model XYC100B-SPT/XYC100B-PRO
should assure that it is used in such an environment.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Separation Distance According to the Frequency of Transmitter (m)
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
Rated Maximum
Output Power of
Transmitter (W)
150kHz to 80MHz
80MHz to 800MHz 800MHz to 2.5GHz
d=1.17 √P
d=1.17 √P
d=2.33 √P
Emissions Test
Compliance
Electromagnetic Environment - Guidance
The Model XYC100B-SPT/XYC100B-PRO uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference to nearby electronic equipment.
RF Emissions, CISPR 11
RF Emissions, CISPR 11
Harmonic Emissions,
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions, IEC 61000-3-3
Group 1
Class B
Class A
Complies
The Model XYC100B-SPT/XYC100B-PRO is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.