Inova Labs XYC100B-SPT User Manual
Page 14

3. It takes up to 3 hours for a fully discharged External Battery to reach a
100 percent charge level when plugged into the unit and an AC or DC
Adapter/Charger. Press the Battery Check button to check the charging
progress.
NOTE: By design, the External Battery will power the POC when
plugged in. Once the External Battery depletes, the unit will switch to
internal battery power. If both the Internal and External batteries are fully
discharged, it can take up to 7.5 hours to achieve a 100 percent charge
level in both batteries.
4. When the External Battery is fully charged, the EXT Battery Check LED
will turn off.
NOTE: The charge indicator will show 4 solid green lights when the
battery has a charge level of 96 percent or more. When the battery
reaches a full 100 percent charge, the EXT LED will turn off.
In the event the power supply is interrupted to the POC when the AC or DC
Adapter/Charger is connected, a two-second audible alarm will sound and
the compressors will transition from external to internal power. During this
transition, you may hear the compressors momentarily pause. No action
is required as the system will automatically transition to internal power and
resume operation.
NOTE: During and after charging, the internal fans may continue to operate
for an additional period of time until the temperature inside the unit is
sufficiently cooled.
REPRESSURIZATION
If the unit has not been used for a period of one or more days, it will
automatically re-pressurize itself. During this time, you may hear the
compressors running for a few seconds.
12
OPERATING
INSTRUCTIONS
continued
* Hours are approximate and based on breaths per minute. As batteries age, charge times and run times
may change.
BATTERY LIFE TIMETABLE
Battery Type
Pulse Setting
1 LPMeq
2 LPMeq
3 LPMeq
Internal Battery*
Sport Model
5.5 Hours
4 Hours
3.25 Hours
Pro Model
12.5 Hours
7.25 Hours
6.25 Hours
External Battery* All Models
3 Hours
3 Hours
2 Hours