Fault

Possible cause

Remedy

❑ No digital pressure read-

ing, warning triangle is

flashing, five or seven

blips*.

Five blips: External sensor

defective or removed?

Seven blips: Internal sensor

defective?

Plug in sensor or use new

one or reconfigure without

sensor.

Contact local distributor.

❑ Warning triangle and

pump symbol are flash-

ing, six blips*.

➨ NT VARIO / VARIO-B pump

and VMS** connected?

✔ Disconnect one component,

restart controller.

➨ Fault at the VARIO pump or at

the VMS?

✔ Check pump and/or VMS or

reconfigure.

➨ Fault at the NT VARIO /

VARIO-B pump?

✔ Check pump and cable con-

nections, restart controller.

❑ Clock symbol is flashing,

control has stopped.

➨ Preselected process time has

passed?

✔ Confirm by pressing key

START/STOP.

❑ Internal venting valve

does not react, valve

symbol is displayed.

➨ External pressure transducer

connected and active?

✔ Select internal pressure

transducer or connect exter-

nal venting valve.

❑ Venting valve does not

respond, valve symbol is

displayed.

➨ Venting valve contaminated? ✔ Clean valve.

❑ Function ”Vac control”:

Control stops, ”arrow up”

is flashing.

➨ Preset maximum pressure

exceeded?

✔ Confirm by pressing key

START/STOP. Adapt maxi-

mum pressure value if nec-

essary.

❑ Function ”Pump down”:

Control stops, ”arrow

down” is flashing.

➨ Pressure below preset mini-

mum pressure?

✔ Confirm by pressing key

START/STOP. Adapt mini-

mum pressure value if nec-

essary.

❑ Function is not displayed,

no menu.

➨ No controllable device con-

nected (valve, VMS, VARIO

pump)?

✔ Connect devices or use con-

troller as measuring device.

❑ Controller does not

respond when pressing

keys (except ON/OFF).

PC symbol is displayed.

➨ Controller in remote mode?

✔ Control CVC 3000 via in-

terface or switch off remote

mode.

❑ Controller does not

respond when operat-

ing any keys. No change

after switching off/on.

✔ Contact local distributor.

❑ Pump does not start or

stops immediately. Warn-

ing triangle and pump

symbol are flashing.

➨ Pump has been exposed to

condensate?

✔ Allow pump to run for some

minutes with atmospheric

pressure at the inlet.

➨ Device fuse blown?

✔ Identify cause of failure. Re-

place device fuse.

➨ Overpressure in the exhaust

line?

✔ Open exhaust line, ensure a

free (pressureless) outlet line.