Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

00:00:00

Vac control

Set vacuum

100 mbar

Hysteresis Auto

Maximum Off

Delay

Off

Duration Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

The screen-shots show the factory-set values.

When selecting ”Graphic” the display shows a pressure over time curve.

Adjustment of the set vacuum during vacuum control:

Dynamic, interactive adaption:

Press the selection knob and keep pressed.

Turning the knob for a 1/4 turn to the left causes pump down.

Turning the knob for a 1/4 turn to the right causes venting.

When the knob is released the actual value is taken over as new set value.

Alternatively:

Fine tuning:

The set vacuum can be adapted only by turning the selection knob

while process is running.

Turn the selection knob.

A full turn causes a change of the set vacuum of 12 mbar.

Turning the knob for one detent causes a change of the set vacuum of 1 mbar.

The timeline in the diagram adapts automatically to the process time.

Press the selection knob twice to return to the standard display.

Vac control

80 mbar

00:13:00