Automatic mode

Switching on the automatic mode
➨ Press key MODE.

☞ ”Auto” is displayed.

☞ After starting process control, the pressure is adapted to

the process automatically: The controller determines the
boiling point and adapts if the boiling pressure changes.

☞ If automatic mode is switched on while process control is

active, the process pressure is adapted.

Setting the maximum speed:
Set the value for the speed by using key

▲ or ▼ (effective

only if the process pressure is adapted):
☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous press-

ing:

☞ Set speed. The new set point is stored when the key is re-

leased.

☞ The reduced maximum speed restricts the evaporation rate

in case of low boiling solvents and a carry over of sub-
stances.

☞ The utilisation of a condenser or a cold trap at the inlet can

be optimised.

If selecting ”HI” (recommended) for motor speed, the

maximum speed for automatic adaptation of the motor
speed is restricted to a value corresponding to the sys-
tem size.

10.0

MODE

MODE

MODE

If automatic switching off was preselected and the controller
has determined that evaporation is completely finished:
☞ ”Auto” is flashing.

☞ The pump is switched off. If the controller measures a pres-

sure > 1060 mbar, the pressure is reduced to 1060 mbar to
avoid overpressure.

☞ The controller is still active, the control must be finished by

pressing key START/STOP or by venting (the coolant valve
will not be closed until then and until the after-running time
has elapsed).

To use the advantages of the automatic mode optimally, switch
on ”Auto” before starting the process.
☞ Switching on automatic mode: Press key MODE.

☞ The controller determines the operating pressure.

☞ When activating the automatic mode if process control is

active (after having pressed key START/STOP) the actual
pressure is stored as set point and adjusted automatically.

Switching off the automatic mode
➨ Press key MODE.

☞ Switching off the automatic mode is possible at any time.

☞ ”Auto” disappears.

☞ While vacuum control is active the actual pressure is stored

as new set point.