Manual determination of the set point p (e. g. boiling point)

➨ Setting the value for the speed by using key ▲ or ▼ (factory

set: 60 Hz).

☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous press-

ing:

☞ Set speed. The new set point is stored when the key is re-

leased.

To start process control:
➨ Start: Press key START/STOP and immediately key p▼.

Keep key p

▼ pressed. Pumping down starts with prese-

lected speed.

☞ Observe the process. When sufficient evaporation occurs,

release key p

▼.

☞ The actual pressure is stored as new set point.

☞ Vacuum control starts.

☞ If the automatic mode is activated additionally (key MODE),

the set point is adjusted automatically in case the boiling
point is drifting.

Setting of set point p (e. g. boiling point)

➨ Setting of the set point by using the keys p▲ or p▼ (factory

set: 100 mbar):

☞ To activate set mode: Press key shortly. ”Set p” appears.

☞ Short tips toggle p in steps of 1 mbar

☞ Continuous pressing activates ramp: Set-point changes with

increasing speed.

Adjusting the process pressure by pumping down:
➨ Press key p▼.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous press-

ing: Pumping down with preset speed, the actual pressure
is displayed and stored as new set point p when the key is
released.

☞ The achieved pressure is stored as new set point and ad-

justed automatically (only in automatic mode).

Adjusting the process pressure upwards
➨ Press key p▲.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous press-

ing: The venting valve is opened, if connected, or pressure
increases due to system parameters. The current pressure
is displayed.

☞ The achieved pressure is stored as new set point when the

key is released and adjusted automatically (only in auto-
matic mode).

p

267

mbar

67

mbar

p

520

mbar

p

Adjusting of set point p during process control

45.0

START

STOP

p