VACUUBRAND MZ 2C VARIO User Manual

Page 25

background image

page 25 of 51

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999034 / 22/04/2008

Manual determination of the set point p (e. g. boiling point)

➨ Setting the value for the speed by using key Mode ▲ or ▼.

☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous pressing:

☞ Set speed. The new set point is stored when the key is re-

leased.

To start process control:
➨ Start: Press key START/STOP and immediately key p▼.

Keep key p

▼ pressed. Pumping down starts with prese-

lected speed.

☞ Supervise process permanently. When sufficient evapora-

tion appears, release key p

▼.

☞ The actual pressure is stored as new set point.

☞ Vacuum control starts.

☞ If the automatic mode is activated additionally (key MODE)

the set point is adjusted automatically in case of change of
boiling point.

Setting of set point p (e. g. boiling point)

➨ Setting the set point by using the keys ps or pt (factory set

25 mbar):

☞ To activate set mode: Press key shortly. “Set p“ appears.

☞ Short tips toggle p in steps of 1 mbar.

☞ Continuous pressing activates ramp: Set-point changes with

increasing speed.

Adjusting the process pressure by pumping down:
➨ Press key p▼.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

Pumping down with preset speed, the actual pressure is dis-
played and stored as new set point p when the key is re-
leased.

☞ The achieved pressure is stored as new set point and ad-

justed automatically.

Adjusting the process pressure upwards
➨ Press key p▲.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

The venting valve is opened if connected or pressure in-
crease due to system parameters, the current pressure is
displayed.

☞ The achieved pressure is stored as new set point when the

key is released and adjusted automatically (only in auto-
matic mode).

p

267

mbar

67

mbar

p

520

mbar

p

Adjusting of set point p during vacuum control

45.0

START

STOP

p