beautypg.com

Vi. set-up – Kistler-Morse KM Load Stand II User Manual

Page 21

background image

17

www.kistlermorse.com

97-1100-01 Rev. H

vI. SET-uP

inStaLLinG a Sun ShieLD

The sun shield reduces sun-induced stresses in the Load Stand sensors and provides additional protection for
the sensors.

1. With the junction box cover off, slightly loosen the screws attaching the junction box to the Load Stand.

2. Slightly loosen the horizontal screw(s) on the bottom flange of the Load Stand.

3. Wrap the sun shield around the Load Stand, slipping the cutout slots behind the loosened screws.

4. Tighten the junction box screws and the horizontal screw(s) on the bottom flange.

5. Replace the junction box cover.

Screw not

on all models

Slide sun shield behind junction
box and loosened screws.
Wrap sun shield around 
Load Stand. Slide shield behind
junction box and loosened screws
on other end.

Loosen Junction Box screws;
Re-tighten after Sun Shield
is installed.

Loosen Bottom Flange Screw(s);
Re-tighten after Sun Shield 
is installed.

caLiBration

There are two calibration methods:

• Live Load calibration — set low span and high span while moving material into or out of the vessel. This is

the preferred method.

• Manual calibration — set scale factor counts, scale factor weight, and zero calibration value without

moving material.

Live Load calibration requires you to move a known quantity of material into or out of the vessel while performing
the procedure. The quantity of material moved must be at least 25% of the vessel’s total capacity to provide best
accuracy. Live Load calibration is also based on the material weight currently in the vessel. Manual calibration
allows you to start using the system as soon as Load Stands and signal processor are installed and wired, even
if you cannot move any (or enough) material now. Manual calibration values are based on system parameters,
including sensor sensitivity, rated load, and signal processor A/D converter sensitivity. These values are known,
can be calculated, or can be obtained from the signal processor. Manual calibration is also based on the material
weight currently in the vessel.

Note that manual calibration does not take into account the actual response to changes in weight. Theoretically,
a change in weight results in a proportional change in digital counts. However, the structure’s actual response
to weight and interaction with piping catwalks, roof, discharge chutes, etc. prevents the system from achieving
theoretical values. Manual calibration is a good start, but to obtain the highest accuracy, perform a Live Load