Appendix, Appendix a - specifications – Iris Sample Processing StatSpin® ThermoBrite Denaturation and Hybridization User Manual
Page 19

15
Appendix
Appendix A - Specifications
Product No.
TS01
TS02
Model No.
S500-12
S500-24
Capacity
Up to 12 slides
Processing Time
0-100 hours
Number of Programs
40
Ramp Time
37
– 95ºC in less than 3 minutes
Cooling Time
95 - 45ºC in less than 6 minutes
Electrical
120 VAC @ 3.0 A
240 VAC @ 1.6A
Dimensions
Depth 45.1 cm
Width 22.8 cm
Height 13.5 cm
Weight 8.5 kg
Environmental
Indoor use
Altitude up to 2000m
Temperature 15ºC to 40ºC
Maximum relative humidity 80% for temperatures up to 15ºC decreasing
linearly to 50% relative humidity at 40ºC
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Pollution degree 2