Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual

Page 7

background image

3

Symbols and Definitions (cont.)

REF

Product/Reference
Number

Indicates the StatSpin product/catalog number

Caution

Statement of caution/warning, read instruction
carefully

Temperature
limitation

Indicates storage requirements range

Do not re-use

Indicates product is single use only

EC Representative

European Community Authorized Representative

IVD

For in vitro
diagnostic use

Clarifies for use as in vitro diagnostic only

Non sterile

Indicates non-sterile product

LOT

Batch code

Indicates product lot number

S N

Serial Number

Indicates instrument serial number code

Use by

Indicates product expiration date

Consult
Instructions

Consult instruction manual or insert sheet for further
explanation

Audible Indicators

Normal Function Codes

Sound

Meaning

2 medium beeps

Sounds on power up; instrument ready

3 short beeps

Cycle completed per specified operating parameters.

Chirp

Start or stop command recognized by the
microprocessor.

Malfunction Codes
1 long beep followed by: 1 short beep

Rotor failed to reach selected rpm within allotted time

1 long beep followed by: 2 short beeps

Lid opened during operation

1 long beep followed by: 3 short beeps

Internal temperature exceeded 60°C (unit can not be

restarted until internal temperature drops below 52°C)

1 long beep followed by: 4 short beeps

Insufficient power to maintain rpm

1 long beep followed by: 5 short beeps

Over current detector tripped.

Cell and Filter Concentrators

A variety of concentration options are available for use with the CytoFuge 2. See the package
inserts for the Filter Concentrators (both reusable and disposable) and the Cell Concentrators (both
reusable and disposable) for detailed information.