Appendix – Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual
Page 14

10
Appendix
Appendix A - Specifications
Product No.
CF02
Model No.
M801-22
Speeds and Forces
600 rpm / 20 xg
700 rpm / 27 xg
850 rpm / 40 xg
1000 rpm / 55 xg
1300 rpm / 93 xg
1600 rpm / 140 xg
2200 rpm / 265 xg
4400 rpm / 1060 xg
Cycle times
2, 4, 6, 8, 10, and 30 minutes
Electrical
24 Volts, DC, 1.7 amp. Includes switching power supply for 100-240
VAC, 50/60 Hz
Diameter 6.8”/ 16.3 cm
Height 5.3”/12.0 cm
Dimensions
Weight 4.75 lbs/2.5 kg
Indoor use
Altitude up to 2000m
Temperature 5ºC to 40ºC
Maximum relative humidity 80% for temperatures up to 31ºC
decreasing linearly to 50% relative humidity at 40ºC
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Environmental
Pollution degree 2