beautypg.com

Myron L ARH1 User Manual

Page 11

background image

X.

TROUBLESHOOTING CHART

17

Symptom

Possible Cause

Corrective Action

No display, even though

Battery weak or not connected.

Check connections or replace battery

measurement key pressed.

(ref. Battery Replacement, pg. 14).

Inaccurate pH readings

1. pH calibration needed (ref. pH Cal., pg. 11).

1. Recalibrate instrument.

2. Cross-contamination from residual

2. Thoroughly rinse sensor well.

pH buffers or samples in sensor well.

3. Recalibrate using fresh buffers (ref.

3. Calibration with expired pH buffers.

pH Buffer Solution, pg. 18).

No response to pH changes

Sensor bulb is cracked or an electro-

Replace pH sensor (ref. pH Sensor, pg. 15).

mechanical short caused by an internal crack.

Will not adjust down to pH 7.

pH sensor has lost KCl.

Clean and rejuvenate sensor (ref. pH, pg. 15)
and recalibrate. If no improvement, replace
pH sensor (ref. pH Sensor Replacement, pg.
15).

pH readings drift or respond

1. Temporary condition due to “memory” of

Clean and rejuvenate sensor (ref. pH, pg. 15)

slowly to changes in

solution in pH sensor well for long periods.

and recalibrate. If no improvement, replace

buffers/samples.

2. Bulb dirty or dried out.

pH sensor (ref. pH Sensor Replacement, pg.

3. Reference junction clogged or coated.

15).

Unstable Conductivity or

Film or deposits on electrodes.

Clean cell cup and electrodes

T D S readings.

(ref. Conductivity or TDS, pg. 15).

Unable to calibrate

Film or deposits on electrodes.

Clean cell cup and electrodes (ref.

Conductivity or T D S .

Conductivity or TDS, pg. 15).