Drain installation – Nor-Lake Stability Chambers User Manual

Page 3

background image

Duplex/ or European Outlet

- (Optional) is located near the top of the interior cabinet back. This

is a 15 amp 115 volt duplex PN 88010300 with its own power supply cord. Duplex outlet power

cord is wrapped up on the back of the cabinet top. Outlet is not inter-wired thru main cabinet

supply and must be plugged into a 115 volt/ or 220/240V power supply. Locate cabinet within 8

feet of the wall receptacle.


Access Port

- (Optional) is provided with a spring loaded cover on the right hand side of the

cabinet.

HUMIDITY INSTALLATION (if applicable)

The water supply should have a dedicated shutoff valve and regulator installed. Connect the water

supply to the humidifier (the operating pressure range is 10-60 psi). For best results, provide a

source of reverse osmosis water with a resistance of .05 to 1.0mΩ, which will reduce the scale

build up and cleaning frequency. Unit is equipped for water supply with 1/8” Male Pipe Threads.

The humidifier reservoir water level may be adjusted by using the adjusting thumb screw located

on the float valve (see below).

The humidifier is a precision built instrument that,

given proper care, will provide years of

dependable service. Cleaning the humidifier

periodically is all that is required to insure proper

operation. Depending on local water conditions,

mineral solids and other matter may accumulate

in various parts of the unit. These accumulations

must be removed so that water flows freely for

efficient operation. Failure to clean the humidifier

may result in malfunction which will lead to repairs.

- 3 -

If higher humidity levels are unobtainable, loosen the

reducing coupling clamp and remove reducing coupling

boot and reducer PVC.


DRAIN INSTALLATION

The Stability Chamber has a floor drain in the bottom of the

cabinet. A silicone drain stopper has been placed in this drain.

The drain stopper will prevent undesirable liquids from running

into the drain. The stopper can be removed if any possible

liquids flowing into this drain are acceptable to go to the floor drain.