beautypg.com

Posey Deluxe Limb Holders User Manual

Application instructions, Rx only limb holders

background image

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

MDSS GmbH

Schiffgraben 41

D-30175 Hannover, Germany

EC REP

APPLICATION

INSTRUCTIONS

1. Feed the strap at the narrow end of the

triangular cuff through the horizontal

slot in the wide end of the cuff. Continue

threading the strap through the loop

located on the outside of the device (fig. 1).

2. Thread the strap from the wide end of

the cuff through the loop in the opposite

direction (fig. 2). For faster application,

the restraint can be attached to the bed

before placing on the patient’s wrist.

1. Use method a. or b. below to attach straps to the bed (repeat

steps 1-5 on each side):

a. Triangulation process; to restrict

patient’s range of motion: Separate the

straps and attach at different points

along a movable part of the bed

frame, out of the patient’s reach, using

quick-release ties (fig. 3).

b. To increase patient’s range of motion: Place the straps together

and attach to a single point along a movable part of the bed

frame, out of the patient’s reach, using quick-release ties.

2. Position the patient’s wrist into the

prepared limb holder. Secure the hook and

loop fastener.

3. Pull the strap on the narrow end of the

triangle snug (fig. 4). Secure the hook and

loop fastener.

4. Slide ONE finger (flat) between the cuff

and the inside of the patient’s wrist to

ensure proper fit. The strap must be snug,

but not compromise circulation (fig. 5).

Before use, check device for damage. Discard if you have

any questions about patient safety.

TO PREPARE LIMB HOLDERS:

APPLICATION INSTRUCTIONS:

• Do not tie a knot at the loop. The knot

would be difficult to untie in an

emergency.

Applicable Products:

REF

2625, 2631

DESCRIPTION OF PRODUCT: Limb holder for limiting limb movement. For bed use only.

Rx ONLY

LIMB HOLDERS

© 2009 Posey Company. All rights reserved.

I9214C 062209

2631

POSEY LIMB HOLDERS

REF

2625 Synthetic Fur Lined

REF

2631 Cotton Flannel Lined

FIG. 5

FIG. 1

FIG. 2

FIG. 3

FIG. 4

INDICATIONS FOR USE:

• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling).
• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital signs.
• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition, causes self-injury, or

compromises wound site integrity.

ADVERSE REACTIONS

• Severe emotional, psychological, or physical problems may occur: if the applied device is uncomfortable;

or if it severely limits movement. If symptoms of these problems ever appear for any reason, get help

from a qualified medical authority and find a less restrictive, product or intervention.

CONTRAINDICATIONS:

DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or suicidal.

MONITOR PER FACILITY POLICY. Check to ensure that:

• Straps cannot slide in any direction or loosen if the patient

pulls on them, or if the bed is adjusted;

• Cuffs are properly secured. If applied too tightly,

circulation will be restricted; if applied too loosely, the

patient may be able to slip his or her limb from the device;

• Cuffs are attached in a way that the patient is not able to

use his or her teeth or otherwise remove the device;

• Cuffs are intact, and not torn or damaged. DO NOT allow

patients to ingest product material.

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS

MAY BE NEEDED:

• If the patient pulls violently against the bed straps;

• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and

down causing self-injury.

See Posey Catalog.

BED SAFETY

• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.

gov/cdrh/beds/modguide.html)

compliant side rails in the UP position

and fill ALL gaps to reduce the risk of

entrapment.

• Use side rail covers and gap protectors to help prevent the

patient’s body from going under, around, through or

between the side rails. A failure to do so may result in

serious injury or death if a patient becomes suspended or

entrapped. Posey offers a full range of side rail pads and/

or gap protectors

to cover gaps.

ADDITIONAL WARNINGS:

5. Adjust the bed strap(s) to allow

desired freedom of movement, without

compromising patient or caregiver safety.

NOTE: To remove cuffs, pull up on the hook

and loop strap to loosen (fig. 6).

FIG. 6

NEVER use on a patient:

With a dislocation or fracture on the restrained limb; or

If an IV or wound site could be compromised by the device.

See the Posey Catalog for other options for such a patient.

ADDITIONAL SAFETY AND LAUNDERING

INSTRUCTIONS ON OTHER SIDE

MADE IN CHINA