Posey EZ-On Mitts User Manual

Application instructions, Rx only finger control mitts

background image

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

MDSS GmbH

Schiffgraben 41

D-30175 Hannover, Germany

EC REP

ApplicAtion

instructions

Applicable Products:

REF

2817, 2826

DEscription oF proDuct: A full padded hand mitt with three finger separators.

rx onlY

FinGEr control Mitts

© 2009 Posey Company. All rights reserved.

I9293 060609

2817

2826

posEY FinGEr control

Mitts

REF

2817 EZ-On Mitts

REF

2826 Economy Mitts

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS

MAY BE NEEDED (See Posey Catalog):
• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and

down and causing self-injury.

MONITOR PER FACILITY POLICY. Check to ensure that:
• Mitts are properly secured. If applied too tightly,

circulation will be restricted; if applied too loosely, the

patient may be able to slip his or her limb from the device;

• Mitts are intact, not torn or damaged, and hook and loop

closes securely. DO NOT allow patients to ingest mitt

material;

• The patient cannot use his or her teeth or otherwise

remove the device and inflict self-injury;

• Monitor closely when the patient is out of bed. Patients

who ambulate while wearing this device may be at risk of

injury from a fall.

ApplicAtion instructions (Repeat steps 1-3 for each mitt):
NOTE: Products are not designed for use with a connecting

strap (See Posey Catalog for other options).

1.

Insert the patient’s hand into the mitt, palm down. Separate

the fingers so that they fit into the three finger separators.

One slot is for the thumb, one is for the little finger, and the

largest slot is for the remaining three middle fingers.

2.

Wrap the wrist strap around the smallest part of the

patient’s wrist, through the plastic ring, and secure with the

hook and loop fastener.

3.

Slide ONE finger (flat) between the device and the inside of

the patient’s wrist to ensure proper fit. The strap must be

snug, but not compromise circulation.

inDicAtions For usE:
• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling).
• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital signs.
• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition, causes self-

injury, or compromises wound site integrity.

contrAinDicAtions:

NEVER use mitts on a patient:

If an IV or wound site could be compromised by the device.

With a dislocation or fracture on the affected limb.

See the Posey Catalog for other options for such a patient.

DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or

suicidal.

ADVErsE rEActions

• Severe emotional, psychological, or physical problems may occur: if the applied device is

uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear

for any reason, get help from a qualified medical authority and find a less restrictive, product or

intervention.

BED SAFETY
• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.

gov/cdrh/beds/modguide.html)

compliant side rails in the UP position

and fill ALL gaps to reduce the risk of entrapment.

• Use side rail covers and gap protectors to help prevent the

patient’s body from going under, around, through or

between the side rails. A failure to do so may result in

serious injury or death if a patient becomes suspended or

entrapped. Posey offers a full range of side rail pads and/

or gap protectors to cover gaps.

ADDitionAl sAFEtY AnD lAunDErinG

instructions on othEr siDE