HT instruments HT32 User Manual
Page 21
HT32
EN - 20
Resistance
Range
Resolution
Accuracy
Open voltage
Overload protection
600.0
0.1
(0.7rdg+2dgt)
1.3V
600V rms
6.000k
0.001k
60.00k
0.01k
600.0k
0.1k
6.000M
0.001M
(1.0rdg+2dgt)
60.00M
0.01M
(1.5rdg+2dgt)
Diode test
Range
Resolution
Accuracy
(0.4 0.8V)
Test
current
Open
voltage
Overload protection
10mV
(1.5rdg+5dgt)
1.5mA
<3V
600V rms
Continuity test
Range
Buzzer
Overload protection
R<500
600V rms
Capacitance
Range
Resolution
Accuracy
Overload protection
6.000nF
0.001nF
(1.9rdg+8dgt)
600V rms
60.00nF
0.01nF
600.0nF
0.1nF
6.000F
0.001F
60.00F
0.01F
600.0F
0.1F
6.000mF
0.001mF
6.1.1. Electrical specifications
Conversion:
TRMS
Measuring rate:
1.5 times per second
Temperature coefficient:
0.15×(accuracy)/°C (<18°C and >28°C)
NMRR Normal Mode Rejection Ratio:
> 50dB for DC parameters and 50/60Hz
CMRR Common Mode Rejection Ratio: >100dB from DC up to 60Hz on DCV
> 60dB from DC up to 60Hz on ACV
6.1.2. Safety
The instrument complies with:
EN 61010-1
Insulation:
class 2, double insulation
Pollution degree:
2
Overvoltage category:
CAT IV 600V, CAT III 1000V (V//A)
CAT IV 500V (A)
For indoor use, max height: