Rounding, Etail – Boltek EFM-100 v.1 User Manual
Page 22
C H A P T E R 2 - I N S T A L L A T I O N
16
Grounding Detail
The EFM-100 field mill needs to be grounded for proper operation and for lightning and
electrical safety. Connect the connector end of the green ground wire to the field mill
ground connector. Connect the other end of the green ground wire and the field mill
mounting hardware to your building’s central ground through a #6 bare ground wire or
other suitable wire according to your local electrical code. Grounding wire and clamps are
available at your local hardware store or electric supplier.
IMPORTANT
The EFM-100 field mill needs to be grounded for proper operation and for
lightning and electrical safety. Connect the field mill green ground wire
and the field mill mounting hardware to your building central ground
through a suitable sized grounding wire according to your local electrical
code.