Pecifications – Perma Pure Baldwin-Series M325 User Manual
Page 3

Section A: Specifications
3
A:
S
PECIFICATIONS
Physical Description
Single (Series) / Dual (Parallel) Channel System
2 x 5” Heat Exchangers connected in series or parallel
1 Passive (cooled to ambient), 1 Active (cooled to 4
°
C) Heat Exchangers
LCD temperature display
Operating Specifications
Sample Gas Flow Range
2-4 LPM
4.3-8.6 SCFH
Inlet Dew Point at Rated
Flow
168
°
F @ 40% H
2
O, 4 LPM
76
°
C
Maximum Cooling Rate
222 BTU/Hr
234kJ/Hr
Dimensions
11.20 x 7.25 x 11.20 in. HWD
28.5 x 18.4 x 28.5 cm
Weight 17
lbs
7.7 kg
Maximum Inlet Sample
Temperature
392
°
F
(200
°
C)
SS, Glass
Impingers
280
°
F
(138
°
C)
Kynar Impinger
Maximum Inlet Pressure
45 psig
3 bar / 2250 mmHg
Maximum Heat Exchanger
Pressure Drop
<+1 in. H
2
O
Ambient Temperature
Range
33-104
°
F
0.56-40
°
C
Outlet Sample Gas Dew
Point
41
°
F
5
°
C
Inlet Tubing Connection
⅜ in. FPT
Outlet Tubing Connection
¼ in. FPT
Drain Tubing Connection
⅜ in. FPT
Voltage
110 (220 optional) VAC
50/60 Hz
Thermoelectric Elements
40 mm
Power Supply
100W
Cooling Down Time
Less than 3 minutes
20% H2O @ 3LPM