Perma Pure Modular Board User Manual
Page 16

Water Slip
UJS
1, 2 or 3
Modbus ID
I d
1 to 255
Set Point 1 is the temperature the cooler will try to maintain impinger #1 at during
operation.
Ready 1 is the temperature that impinger #1 must be below to turn on the relay #1
and external equipment (i.e. sample pump)
Set Point 2 is the temperature the cooler will try to maintain impinger #1 at during
operation.
Ready 2 is the temperature that impinger #2 must be below to turn on the relay #2
and external equipment (i.e. sample pump)
Display set the method to control the channel display of data. Toggle, shown as
“toG”, will display each channel for 10 seconds constantly switch between the
two. Switch, shown as “SUJ”, will allow the user to select the channel displayed
through an external switch.
Water Slip set the relay used to alarm for water present at the Water Slip sensor.
Setting of 1 will have the water slip sensor control relay #1 with ready for channel
#1. Setting of 2 will have the water slip sensor control relay #2 with ready for
channel #2. Setting of 3 will have the water slip sensor control relay #2 alone; the
channel 2 ready alarm is ignored.
Modbus ID is the address used in serial communication to the unit.
NJ Thermocouple Display
DIP Switch 1, 2 and 3 are ON, ON and OFF.
Menu
Item
Display Range
Ready
1 rd1
-10
to
25
°C
Ready
2 rd2
-10
to
25
°C
Display dSP
Toggle
or
Switch
Water Slip
UJS
1, 2 or 3
Ready 1 is the temperature that impinger #1 must be below to turn on the relay #1
and external equipment (i.e. sample pump)
Ready 2 is the temperature that impinger #2 must be below to turn on the relay #2
and external equipment (i.e. sample pump)