Odisi b – Luna ODiSI B User Manual
Page 2
Designs and specifications are subject to change without notice.
Please refer t
ODiSI B
LTODiSIB10 REV. 007 03/03/2013
©Luna Innovations Incorporated, Roanoke, VA All rights reserved.
PARAMETER
SPECIFICATION
UNITS
Performance
Standoff
50
meters
Wavelength Accuracy
1,2
1.5
pm
Strain Range
± 10,000
µStrain
Temperature Range
3
-50 to 300
°C
Mode of Operation
4
Standard
High-
Speed
High-
Resolution
Extended
Length
Maximum Sensing Length
10
2
10
20
meters
Acquisition Rate
100
250
23.8
50
Hz
Sensor Spacing
5
5
1.25
5
mm
Gage Length
5
5
1.25
5
mm
Strain Repeatability
(Single-scan)
± 5
± 5
± 20
± 10
µStrain
Temperature Repeatability
(Single-scan)
± 0.4
± 0.4
± 1.6
± 0.8
°C
Electrical Trigger
Acquisition
TTL compatible
Event
TTL compatible
Physical
Dimensions
14 x 12.5 x 6.75
in
Weight
17.3
lb
Power Consumption
50
W
CLASS 1 LASER PRODUCT
1 Calibration is performed internally using a NIST-traceable HCN gas cell
2 Temperature and strain accuracies from spectral shift of Rayleigh scatter are 0.15 °C and 1.25 µStrain, calculated
using the default conversion coefficients 1 GHz = 0.8 °C = 6.58 µStrain [Othonos and K Kalli, Fiber Bragg Gratings
(Artech House, Boston, 1999)].
3
Based on material properties of the standard sensor: polyimide-coated, low-bend-loss optical fiber. For
temperatures outside of this range, please contact your Luna representative for more information.
4
Base configuration includes one mode of operation. Additional modes are upgrade options.