AVSL 600.100 MTM01 User Manual
Page 10
MTM01 User Manual
CAPACITY MEASUREMENT
1.
C
onnect the BLACK test lead to the COM jack and the RED to
the mA jack.
2. Set the Range switch at F position. (NOTE: The polarity of the
RED lead is positive “+”)
3. Connect test leads across the capacitor under measure and be
sure the polarity of connection is observed.
Please Note:
To avoid damage to the Meter, disconnect circuit power and
discharge all high-voltage capacitors before measuring
capacitance. The tested capacitor should be discharged before the
testing procedure. Never apply voltage to the input, or serious
damage may result.
CONTINUITY TEST
1. Connect the BLACK test lead to the “COM” jack and the RED
to the “VΩ” jack (Note: The polarity of the red test lead is
positive “+”).
2. Set the range switch to range
3. Connect the test leads across the load to be measured.
4. If the circuit resistance is lower than about 30±20Ω, the built-
in buzzer will sound.
DIODE MEASUREMENT
1. Connect red lead to “VΩmA”, black lead to “COM”.
2. Set RANGE switch to “
” position.
3. Connect the red test lead to the anode of the diode to be
measured and black test lead to cathode.