1.4 Power Supply Unit Front Panel Layout

The power supply front panel contains the input ON/OFF circuit breaker

and the FAULT, MASTER / STANDALONE and SLAVE indicators. Refer

to Figure 1-2 and Table 1-3. for a description of front panel controls and

indicators.

Figure 1-2.

Model 4Q06125PS Front Panel

Table 1-3.

Power Supply Front Panel Controls and Indicators

Reference

(Figure Fig-

ure 1-2.)

Control or Indi-

cator

Function

1

POWER ON/OFF

circuit breaker

switch

Applies source power to unit

2

SLAVE indicator

Not used for single power supply configuration of Model
4Q06125PS

3

MASTER indicator

Lights when configured as Model 4Q06125PS single
supply

4

FAULT indicator

Lights red when a fault is detected. The following failure
or fault conditions can cause the FAULT indicator to light:
overtemperature, instant internal overcurrent, output
overvoltage/overcurrent, local ±15V failure, input under/
overvoltage, input overcurrent, internal output under/
overvoltage, internal output overcurrent, overtempera-
ture, fan failure.
When the FAULT indicator lights, an audible beep
sounds a warning for approximately two seconds and the
output is crowbarred by an internal contactor. The fault is
latched. After the cause of the fault is removed, the unit
can be restarted by cycling the POWER circuit breaker to
OFF, then ON, or by applying a START_EXT pulse at
Analog I/O Port pin 7