2nd Ave. BF-40 User Manual
Bronchoscope disinfection guide, For oes, Bf-40 series bronchofiberscopes
Alcohol
To Suction Pump
To Suction Pump
Suction Cleaning Adapter
Single-Use Suction Valve (MAJ-209)
Suction Cylinder
Single-Use Biopsy Valve (MAJ-210)
Universal Cord
Instrument/Suction Channel Port
Distal Tip
Bending Section
Light Guide
Boot
UP/DOWN Angulation Control Lever
Control Section
Insertion Tube
Venting Connector
HIGH-LEVEL DISINFECTION
RINSING
TRANSPORTATION
DRYING
NOMENCLATURE
IMPORTANT
1
2
3
4
5
6
9
10
7
8
Attach a syringe to the adapter.
Pull the plunger of the syringe to
fill the instrument/suction channels
and channel cleaning adapter
with disinfectant solution.
Fill basin with disinfectant. Attach a
suction cleaning adapter (MAJ-222) to
the previously cleaned bronchoscope.
Immerse completely in disinfectant
solution.
While immersed, detach the adapter
from the bronchoscope. Use a clean,
lint-free cloth to wipe away any bub-
bles clinging to external surfaces.
Soak all equipment according
to the time and temperature
recommended by the disinfectant
manufacturer. Use a timer to mea-
sure the exact contact time.
Connect the suction cleaning adapter
and syringe. Inject air to remove
disinfectant from the instrument/suction
channels, then remove the broncho-
scope from the disinfectant.
Disconnect the suction cleaning
adapter and immerse the broncho-
scope and adapter in sterile water.
Using a sterile, lint-free cloth, thor-
oughly rinse and wipe all surfaces.
Connect suction cleaning adapter
to bronchoscope and to a tube
from a suction pump. Aspirate
water through channels for 30
seconds. Remove from water and
aspirate air through channels for
60 seconds.
Turn the suction pump off. Hold the control
section with the openings pointing down and
disconnect the suction cleaning adapter.
Dry the bronchoscope and adapter with an
alcohol-moistened cloth. Make sure that the
equipment is completely dry prior to storage.
Hold all parts of the endoscope when carrying. Do
not let the ends of the bronchoscope swing freely.
Clean the bronchoscope
meticulously prior to disinfection.
• Wear appropriate personal protective
equipment during reprocessing.
• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically
with all valves and caps removed.
• Use a compatible disinfectant and test it
for proper concentration as recommended
by the manufacturer.
• Do not use the carrying case to store
"patient-ready" endoscopes. Any
bronchoscope placed in the carrying case
must be reprocessed prior to patient use.
Immerse the bronchoscope's distal
end in 70% alcohol and aspirate
alcohol for 5 seconds. Remove the
distal end from the alcohol and
aspirate air for 20 seconds.
©
2
0
0
4
O
ly
m
pu
s
A
m
er
ic
a
In
c.
/
P
ri
nt
ed
in
U
SA
0
1
/0
4
—
P
W
C
-1
0
0
0
2
0
for OES
®
BF-40 Series Bronchofiberscopes
BRONCHOSCOPE DISINFECTION GUIDE
Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING:
This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.