beautypg.com

Maintenance – Invacare IRC1001 User Manual

Page 12

background image

12

MAINTENANCE

WARNING

Electrical shock hazard. DO NOT remove outer case of

this unit. All disassembly and maintenance of this unit

MUST BE done by a qualified Invacare Dealer.

CLEANING THE OUTER CASE

WARNING

UNPLUG unit before cleaning. DO NOT submerge in

water to clean. Clean outer case with damp cloth.

CLEANING THE NEBULIZER

NOTE: With proper maintenance your nebulizer should last 30 days. Since the nebulizer

is disposable, it is recommended that an extra nebulizer be kept at all times.
NOTE: Clean the Nebulizer after every use and at least once a day.
NOTE: Your physician and/or a Invacare Dealer may specify certain cleaning

procedures. If so, follow their recommendations. If not, clean according to the

following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.
2. Disassemble nebulizer chamber by turning counterclockwise and

separating.

3. Fill two (2) plastic containers or bowls:

A. Washing: Hot water/detergent solution.

B. Soaking: Hot water/vinegar solution (one [1] part vinegar to

three [3] parts water).
4. Thoroughly clean the nebulizer and mouthpiece in hot water/

detergent solution.

5. Remove from the water/detergent solution and rinse with clear hot

tap water.

6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.

MAINTENANCE