beautypg.com

Cleaning/replacing the air filters (figure 2), Replacement parts, Accessories – Invacare 50 psi Compressor IRC607 User Manual

Page 3

background image

3

Pressure

Gauge

Outlet Port

ON/OFF

Switch

Power Cord

Inlet Filter

Assembly

Pressure

Control

Knob

FIGURE 2 - OPERATION

4. Turn the pressure control knob fully counterclock-

wise.

NOTE: As the knob is turned counterclockwise the pres-
sure will decrease, as the knob is turned clockwise the
pressure will increase.

5. Attach the nebulizer tubing to the outlet port.

6. Turn the ON/OFF switch to the ON position.

WARNING

Use only the pressure setting directed by your
physician or qualified healthcare professional.

7. Turn the pressure control knob clockwise to adjust

the pressure to the setting specified by your
physician's instructions.

NOTE: The pressure gauge needle will vibrate. Adjust
the pressure gauge setting as closely as possible to the
specified pressure.

8. Proceed with therapy per physician's instructions.

CLEANING/REPLACING THE AIR
FILTERS (FIGURE 2)

NOTE: Check the filters at least once each month.
If the compressor begins to lose pressure in nor-
mal use, check the air filters.

1. Remove the filter cap from the Inlet Filter Assembly.

2. If the foam filter appears dirty, wash with mild

detergent and water. Rinse thoroughly and let
air dry.

WARNING

DO NOT wash the felt filter. Water damages
the felt filter.

3. Replace the felt filter if it appears dark in color,

becomes clogged, or is damaged.

WARNING

DO NOT operate this compressor without fil-
ters installed.

REPLACEMENT PARTS

Felt Filter
Foam Filter
Filter Cap

ACCESSORIES

WARNING

DO NOT use the compressor for continuous op-
eration more than 10 hours, this can result in over-
heating and damage to the compressor.

J-Tube and DISS Connector for large volume
nebulizers

Suction Feet