Part No 1152408

3

Jasmine™ Scale JLS5

Installing the Jasmine Scale

NOTE: For this procedure, refer to FIGURE 2.

ƽ WARNING

Patient and sling MUST be removed from the lift during ALL installation procedures.

1. Position the load cell assembly of the Jasmine Scale into the boom mounting bracket (Detail ʺAʺ of 

FIGURE 2).

2. Secure the Jasmine Scale to the boom mounting bracket with a bolt, washers, bushings, nylon 

washers, locknut and the top rubber cap. Securely tighten. Install the small rubber cap.

NOTE: Refer to Detail ʺAʺ of FIGURE 2 for washer and small rubber cap orientation.

NOTE: Ensure the shaft of the bolt passes through both sides of the boom mounting bracket.

3. Insert swivel bar pin with an o‐ring and a washer through the swivel bar (Detail ʺBʺ of FIGURE 2).

4. Align the mounting holes in the swivel bar pin with the mounting holes in the load cell assembly 

(Detail ʺBʺ of FIGURE 2).

5. Secure swivel bar pin to the load cell assembly with a screw and locknut. Tighten securely (Detail 

ʺBʺ of FIGURE 2).

ƽ WARNING

After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is

tightened securely - otherwise injury or damage may occur.

6. Weigh the patient. Refer to Operating the Scale and Weighing the Patient in the Jasmine Patient Lift 

Owner’s Manual, Invacare P/N 1150704.

FIGURE 2 Installing the Jasmine Scale

Locknut

Swivel Bar

Washer

Swivel Bar Pin

Screw

Jasmine

Scale

Jasmine

Scale

Load Cell
Assembly

Bolt

DETAIL “B”

DETAIL “A”

Locknut

Nylon

Washers

Washer

O-Ring

Top Rubber

Cap

Bushing

Small

Rubber

Cap

Boom

Mounting

Bracket

Bushing

Washer