Installing the jasmine scale, Ƽ warning – Invacare Jasmine JLS5 User Manual
Page 3
Part No 1152408
3
Installing the Jasmine Scale
NOTE: For this procedure, refer to FIGURE 2.
ƽ WARNING
Patient and sling MUST be removed from the lift during ALL installation procedures.
1. Position the load cell assembly of the Jasmine Scale into the boom mounting bracket (Detail ʺAʺ of
FIGURE 2).
2. Secure the Jasmine Scale to the boom mounting bracket with a bolt, washers, bushings, nylon
washers, locknut and the top rubber cap. Securely tighten. Install the small rubber cap.
NOTE: Refer to Detail ʺAʺ of FIGURE 2 for washer and small rubber cap orientation.
NOTE: Ensure the shaft of the bolt passes through both sides of the boom mounting bracket.
3. Insert swivel bar pin with an o‐ring and a washer through the swivel bar (Detail ʺBʺ of FIGURE 2).
4. Align the mounting holes in the swivel bar pin with the mounting holes in the load cell assembly
(Detail ʺBʺ of FIGURE 2).
5. Secure swivel bar pin to the load cell assembly with a screw and locknut. Tighten securely (Detail
ʺBʺ of FIGURE 2).
ƽ WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is
tightened securely - otherwise injury or damage may occur.
6. Weigh the patient. Refer to Operating the Scale and Weighing the Patient in the Jasmine Patient Lift
Owner’s Manual, Invacare P/N 1150704.
FIGURE 2 Installing the Jasmine Scale
Locknut
Swivel Bar
Washer
Swivel Bar Pin
Screw
Jasmine
Scale
Jasmine
Scale
Load Cell
Assembly
Bolt
DETAIL “B”
DETAIL “A”
Locknut
Nylon
Washers
Washer
O-Ring
Top Rubber
Cap
Bushing
Small
Rubber
Cap
Boom
Mounting
Bracket
Bushing
Washer