SECTION 5—MAINTENANCE
25
SECTION 5—MAINTENANCE
ƽ WARNING
Before performing any maintenance to the tub, make certain that the electric power
source is shut off to avoid electrical shock.
Be certain ALL electrical work is performed by a qualified electrician and is in
compliance with local codes.
If any part of the system is NOT functioning properly, DO NOT use the system. Have
the system serviced to correct the problem.
The system MUST be maintained on a regular basis to ensure it is functioning properly.
ELECTRICAL SHOCK HAZARD. All disassembly and maintenance of the tub MUST
be done by a qualified technician, certified electrician or plumber.
Disconnect electrical supply or turn OFF circuit breaker before performing any
maintenance to the tub.
Parts Information
NOTE: For parts information, please contact Invacare Continuing Care Group Technical
Service at (866) 616‐2631.
NOTE: Every six months or as necessary, have a qualified technician perform a thorough inspection and
servicing. Routine maintenance will extend the life and efficiency of your tub.
Safety Inspection Checklists
Initially
❑ Ensure that the maximum water supply temperature does not exceed 110° F (43°C).
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Verify the water flow temperature with a separate thermometer.
❑ Visually inspect for cracks in the tub surface.
❑ Check that all labels are present and legible. Replace if necessary.
Before Each Use
❑ Disinfect the tub.
❑ Ensure that the maximum water supply temperature does not exceed 110° F (43°C).
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Visually inspect for cracks in the tub surface.
Every 80-150 Baths
❑ Perform the heavy duty cleaning procedure on the tub. Refer to Heavy Duty Cleaning