Calibration path – Health O Meter PRO PLUS 2650KL User Manual
Page 21

20
CALIBRATION PATH
The calibration is performed using kg or lb, according to the units of measure used upon
entering into the calibration path.
Operator Action
Display
1. Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
Calibration
Process
2. Press
ENTER
Enter load weight
600.0
3. Using
cd keys and/or the keypad, set the calibration
load to at least 200 lbs (for best results and greater
accuracy, use a 600 lb weight). Press ENTER
Zero calibration
Please clear the scale
4. Please clear the scale platform and press ENTER
Zero calibration
Please wait
5. Please wait without touching or shaking the scale
until the zero calibration process is finished
Weight calibration
Put: XXX.X
6. Load the scale platform with the required weight and
press ENTER
Weight calibration
Please wait
7. Please wait without touching or shaking the scale
until the calibration process is finished
Calibration
factor: X.XXXXX
8. Press
ENTER
Calibration
Please clear the scale
9. Please remove weight from the scale platform
Rebooting
Please wait
10. Please wait until the scale resumes normal operation
Health O Meter Pro Plus
EXPLODED VIEW OF SCALE
Display Assembly