Potential for radio/television interference, Troubleshooting – HoMedics BPA-300 User Manual

Page 16

background image

30

29

POTENTIAL FOR RADIO/TELEVISION
INTERFERENCE

For USA only:
This product has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of FCC rules. These limits are
designed to provide reasonable protection against harmful interference in
a residential installation. The product generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If the product does cause harmful
interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct
the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to

which the receiver is connected.

• Consult the dealer or an experienced radio/TV technician for help.

For Canada only:
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing
equipment standard entitled “Digital Apparatus”, ICES-003 of the
Canadian Department of Communications.

Cet appareil numérique respecte les limites de bruits radioeléctriques
applicables aux appareils numériques de Classe B prescrites dans
la norme sur le matériel brouilleur: “Appareils Numériques”,
ICES-003 édictée par le ministre des communications.

TROUBLESHOOTING

If any abnormality arises during use, please check the following points

Note: If the unit still does not work, contact HoMedics Consumer

Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.

SYMPTOMS

POSSIBLE

CAUSES

CORRECTION

Unit does not turn on when
the START/STOP
button is pushed.

Batteries have run down.

Replace them with four new
AA alkaline batteries.

Battery polarities have been
positioned incorrectly.

Re-insert the batteries in the
correct positions.

EE measurement error
symbol shown on display or
the blood pressure value is
displayed excessively low
(or high).

Cuff has been placed
incorrectly.

Wrap the cuff properly so
that it is positioned
correctly.

Did you talk or move
during measurement?

Measure again. Keep arm
steady during
measurement.

Shaking of the arm with the
cuff on.

Er error symbol shown on
display - measurement
result exceeds
measurement range.

Did you move during
measurement?
Shaking of the arm with
the cuff on.

Measure again. Do not
move during measurement.

E1 error symbol shown on
display

Air circuit abnormality. Cuff
tube may not be plugged
into monitor correctly.

Check cuff connection.
Measure again.

E2 error symbol shown on
display

Inflation pressure exceeding
300 mmHg.

Turn the unit off, then
measure again.

E3 error symbol shown on
display

Data error.

Remove and reload the
batteries.