Hach HCT 159 User Manual

Page 2

background image

ZERO

5

1

3

4.0

ml

Sample

2

4

15

min

PROCEDURE

for Free Iron

applies to DR/4000, DR/2500 DR/2010 and DR/800

Note
If this is the first time you have run this revision of the test on
your DR/4000, you must contact your local Hach distributor
for an upgrade of your software (Cat.No. 49544-00). Software
Version number 2.39 contains the factory UniCell™ programs.

INSTRUMENT SET-UP

1. Press the soft key under HACH PROGRAM (or

USER PROGRAM if the method was entered as a
USER PROGRAM).
Enter program number 2130 (or 159 if the method was
entered as a USER PROGRAM).
Press ENTER.

2. The display will show:

HACH PROGRAM: 2130
Iron, HCT 159
The wavelength (

λ

), 485 nm, is automatically selected.

3. Insert the Test N’Tube adapter into the cell holder by

sliding it under the thumb screw and into the alignment
grooves. Fasten in place by tightening the thumb screw.

4. Follow the procedure steps.

(Note: If measuring Total Iron and performed the sample
digestion using the HCT 200 Metal Prep Set, select the
Total Iron alternate form under the Options soft key menu).

5. Wipe the outside of the zero vial.

Place the zero vial into the adapter.
Close the light shield.

6. Press the soft key under ZERO.

The display will show: 0.00 mg/L Free-Fe or
0.00 mg/L Tot-Fe if the Total Iron alternate form
was selected.

7. Clean the outside of the sample vial.
8. Place the prepared sample vial into the adapter. Close

the light shield. Results in mg/L Free-Fe or Tot-Fe are
displayed.

Note
If the test overranges, repeat the UniCell™ test with diluted
sample.

2. Pipet 4.0 ml sample

into the sample vial.

3. Close sample vial and

invert a few times, until
the freeze-dried
contents are
completely dissolved.

4. Start the instrument

timer.
A 15-minute reaction
period will beginn.

5. Zero the instrument

using the zero vial
(white cap) provided
and read the sample.

DR / 4000—Rev 2.39

1. Enter program number

and set up the
instrument as explained
for each instrument
type.

HCT 159

HCT 159

LIST OF REQUIRED REAGENTS

Iron-Fe
UniCell™ HCT 159

23/pkg

HCT 159

OPTIONAL REAGENTS

Metal Prep Set
UniCell™ HCT 200

50 digestions HCT 200

Iron Standard
1000 mg/L as Fe

100 ml

2271-42

Sodium Hydroxide
Standard Solution

1000 ml

2450-53

(5.0 N)

Nitric Acid, concentrated

500 ml

152-49

OPTIONAL APPARATUS

Cylinder, mixing – 100 ml

each

20886-42

Flask, volumetric 100 ml

each

547-42

Pipettor, (Jencons) 1-5 ml

each

27951-00

Replacement tips for 27951-00

pk/100

27952-00

Pipettor, (Jencons) 100-1000 µl

each

27949-00

Replacement tips for 27949-00

pk/400

27950-00

pH Paper

pk/100

26013-00

MEASURING RANGE:

0.10 - 5.00 mg/L Free-Fe / 0.12 - 6.00 mg/L Tot-Fe

PROCEDURE

for Total Iron:

After digestion with the Metal Prep Set HCT 200,
follow the procedure steps 1 – 5.