Hamilton Electronics EPHUM011 User Manual
Page 28

pH ARC Sensors Modbus RTU Programmer
’
s Manual (EPHUM011)
624300/01
page 28 / 67
2.6.2.1
Description of PA9 (Moving Average)
In register 3360, a plain text ASCII description of PA9 is given.
Start
register
Number of
registers
Reg1
… Reg8
16 ASCII characters
Modbus
function code
Read
access
Write
access
3360
8
Description of PA9
3, 4
U/A/S
none
Figure 2.6.2.1.1: Definition of register 3360.
Command: Moving average text
Modbus address:
3360
Length:
8
Type:
3
Read
Parameter:
Text
Format:
Character
Value:
Moving average
Figure 2.6.2.1.2: Example to read the description for
“
Moving average
“
.
2.6.2.2
Selecting the Physical Unit and Writing the Value for PA9
In register 3368, the available physical units for PA9 are defined.
Start
register
Number of
registers
Reg1 / Reg2
(bitwise defined)
Modbus
function code
Read
access
Write
access
3368
2
Available physical units for PA9
3, 4
U/A/S
none
Figure 2.6.2.2.1: Definition of register 3368.
Command: Moving average av. units
Modbus address:
3368
Length:
2
Type:
3
Read
Parameter:
Units
Format:
Hex
Value:
0x01
Figure 2.6.2.2.2: Example to read the available physical units for PA9. The only one available here is
“
none
”(0x01). For the definition of the physical units see chapter 2.5.1.
Start
register
Number of
registers
Reg1 / Reg2
(bitwise defined)
Reg3 / Reg4
Modbus
function code
Read
access
Write
access
3370
4
Select physical unit
for PA9
Value for PA9
(1-16, default: 2)
16
none
S
Figure 2.6.2.2.3: Definition of register 3370. Only one bit for the physical unit can be set. PA9 can be
set to the value 1-16. A value of 1 does not influence the response time of the sensor, a value of
16 increases the response time of the sensor to 48 s.
By writing to register 3370 the active physical unit for PA9 can be selected by choosing one of the
physical units that are defined in register 3368. The value of the parameter can be set as well.
Command: Moving average
Modbus address:
3370
Length:
4
Type:
16
Write
Parameter:
Unit
Value
Format:
Hex
Decimal
Value:
0x01
12
Figure 2.6.2.2.4: Example to set the physical unit of PA9 to
“
none
”(0x01) and the value of the moving
average to 12.