Sensor Installation: F-Series
21
Purge Fittings
8. Make sure the purge fitting seals are tight so air cannot be drawn by suction into the sensor case.
Table 4
Time required to purge F-Series sensor cases
Sensor model
Purge rate
ft
3
/hr (l/hr)
Time
(1)
minutes
(1) If purge lines are being used, increase purge time to fill the additional volume.
F025
20 (566)
4 1/2
F050
20 (566)
4 1/2
F100
20 (566)
6
F200
20 (566)
15
F300
20 (566)
25