Specifications – Amprobe GP-1 Earth-Tester-Ground-Probe User Manual
Page 18

RESISTANCE
Range
Resolution
0 - 19.99
Ω
0.01
20.0 - 199.9
Ω
0.1
200 - 1999
Ω
1
Accuracy:
+/- (2% of reading + 2 Digits)
RESISTIVITY
Range
Resolution
0-19.99
Ω
m
0,01
20.0-199.9
Ω
m
0.1
200-1999
Ω
m
1
2.00k - 19.9k
Ω
m
10
20.0k - 199.9k
Ω
m
100
200k - 314
Ω
m
1000
Accuracy:
+/- (2% of reading + 2
π
a.0.02
Ω
); where
ρ ≤
19.99
Ω
2
π
a
+/- (2% of reading + 2
π
a.0.2
Ω
); where 19.99
Ω
<
ρ ≤
199.9
Ω
2
π
a
+/- (2% of reading + 2
π
a.2
Ω
); where 19.99
Ω
<
ρ
2
π
a
Test current:
<10mARMS sine wave
Test Frequency:
135 Hz+/- 1Hz
VOLTAGE
Range:
0-600VRMS max
+/- (2% of reading + 2 Digits)
INCLUDED ACCESSORIES
Model GP-CS”
(1) 30’(10m) Green test cable
(1) 30’(10m) Black test cable
(1) 47’(15.5m) Red Test cable
(1) 77’(25.5m) Blue test cable
Model GP-MR
(4) 9”measuring rods
Model GP-TL
(1) Safety banana test lead set
Model GP-RS232
(1) 9 pin male to 9 pin female-RS-232 cable
Model GP-BP
(1) Black shoulder pouch
Download Software
GENERAL
Fuse:
5 x 20mm, 32mA @250V fast blow, Amprobe Model 5 x 20-250.32
Input Power:
(4) “D” cells, Amprobe Model MN-1300
Display:
3-1/2 Digit L.C.D. 3/4”(H)
Weight:
7.26 lbs. (3.3kg)
Dimensions(WxHxL):
13.6”x5.1”x9.8” (3.45Cm x 13Cm x 25cm)
Nominal Temp. Range:
41 - 95°F (5 - 35°C)
Working Temp. Range:
32 - 104°F (0 - 40°C)
Storage Temp. Range:
14 - 140°F (-10 - 60°C)
STANDARDS
Constructed according to:
EN61010-1
Ground resistance measurement:
VDE 0413 Part 7
Radio frequency emission
EN50081-1
Radio frequency immunity
EN50082-1 Class B
Specifications