Amprobe AMB-50 Insulation-Resistance-Tester User Manual
Page 16
![background image](https://www.manualsdir.com/files/864280/content/doc016.png)
10
SPECIFICATIONS
Safety and Compliances
Certification
�
Compliances
IEC 61010 CAT.III 600V overvoltage and double insulation standard
Physical Specifications
Display (LCD)
Digital: 1999 counts
Analog bar graph.
Operating Temperature
0°C~40°C (32°F ~104°F )
Storage Temperature
-20°C ~60°C (-4 °F~152°F )
Relative Humidity
85% @ -10°C ~40°C below;
90% @ 0°C ~40°C :
Battery Type
8pcs X 1.5V (LR14) batteries or power adaptor (input voltage 230V, 50/60Hz,
75mA, input DC14V, 1.0A).
Power adaptor is not included
Dimensions (H x W x L)
202 x 155 x 94 mm
Weight
Approx. 2kg (4.4 Lbs) (including battery)
General Specifications
Range
Auto
Overloading
Display OL on insulation resistance range
Battery Indicator
Display
Icon Display
Equips with function and battery indicator icons.
Current Consumption
Maximum: around 1.0A
Average: around 20mA
Feature Summary
Display Backlight
Bright backlight for clear readings in poorly lighted areas.
Computer connection
Via USB interface.
Data Logging and Recall
18
Autorange
The Meter automatically selects best range
Warning
and red light will on.
Voltage
Auto release voltage
COMP Measurement
Use the Compare function to set a pass/fail compare level for the insulation
measurements.
PI Measurement
Polarization Index is the ratio of insulation resistance. You can pre-set two point
of times and automatically carry out the measurement.
TIME
To carry out measurement by setting a specified time within 15 minutes.