Amprobe ACD-45PQ Power-Quality-Clamp User Manual

Page 18

background image

12

3-Phase 4-Wire (3

4W) Unbalanced-Load Power Function

• Set the slide-switch function-selector to the “3~Un-Bal” Power position.

Press “3W4W” button momentarily to select 4-Wire measurements.
Annunciator “4W” turns on.

• Clamp the jaws around “Line 1” as reminded by annunciators “ L1”,

and connect Black test probe (COM terminal) to “Line n (neutral)” and
Red test probe (+ terminal) to “Line 1” as reminded by annunciators “

” on mini-display.

• When the reading is stable, press “

” button momentarily to enter the

first measuring value.

• Then clamp the jaws around “Line 2” as reminded by annunciators “

L2”, and connect Black test probe (COM terminal) to “Line n (neutral)
and Red test probe (+ terminal) to “Line 2” as reminded by annunciators

” on mini-display.

• When the reading is stable, press “

” button momentarily to enter the

second measuring value.