Amprobe TMD-10 Dual-Temperature-Meter User Manual
Page 8
7
To use the Relative mode for T2, disable
and press and hold
key
> 1 second to swap T2 to the top display and T1 to bottom display.
T1 T2 / T1-T2
T1 T2 displays the 2 inputs independently. Press and hold
button > 1
second to swap display location of the display location of T1 and T2.
(T1-T2)
Press
button to display T1 minus T2 temperature value. The top display
shows the T1-T2 temperature. The lower display shows the top display
selected in the T1 T2 mode. Press
button again to disable this function
Auto Power off (APo)
The TMD-10 powers down automatically after approximately 15 minutes of
inactivity to save battery life. Press the � button to turn the TMD-10 ON. You
can disable Auto Power Off by pressing the
button > 1 second after the
TMD-10 is ON.
�
The power cannot be turned oFF in the oFFSET SETUP mode.
Thermocouple Error Correction (oFS, offset)
User can add an offset to compensate for the error of specific thermocouple
probes.
Press and hold the
1.
(OFS) button > 1 second to enter the Offset Setup
mode.
Immerse the thermocouple(s) in thermos bath with crushed ice and fresh
2.
water (0.0 °C / 32.0 °F).
Press the
3.
button to increase or power � button to decrease the
Offset value. The allowable adjustment range is ± 5 °C / ± 9 °F in 0.1°C /°F
increments.
Press
4.
(SET) button to save and leave setup mode. The “Offset” icon
will remain on the LCD.
Press and hold
5.
button > 1 second to swap T2 to the top display and
T1 to bottom display. Press
button to switch T2 from type K to type J.
Repeat steps 1, 2, 3 and 4 for T2.
6.
To remove the Offset, Repeat steps 1, 2, 3 and 4 setting the offset (lower
7.
display) to 0.0. When both thermocouple offset are zeroed out, the
“Offset” icon will be turned off.