beautypg.com

Amprobe CO2-100 CO2-Meter User Manual

Page 17

background image

11

TROUBLE SHOOTING

Can’t power on

• Make sure you press power key more than 0.3

second.

• Check the battery conditions and replace if

necessary.

• Check whether the adaptor is well plugged.
• Move batteries away for one minute and then

re-install.

Display disappear

• Check whether the low battery icon is appeared

before the display is off. If yes, replace with new

batteries.

Fixed readings
• Check whether data hold function was activated.

(HOLD icon at the left top)

Slow response

• Check whether the air flow channels on the rear

were blocked

Error code

E01: CO2 sensor damaged.
E02: The value is under range.
E03: The value is over range.
E04: The original data error results in this error

(DP, WB)

E07: Too low voltage to measure CO2. Replace

batteries or use an adaptor.

E11: Retry humidity calibration.
E17: Retry CO2 calibration.
E31: Temperature sensor damaged.
E34: Humidity sensor damaged.