Amprobe AM-42 Multimeters User Manual
Page 7

7
DC Voltage (AM-42)
RANGE Accuracy
400.0mV 1.0%+2d
4.000V, 40.00V, 400.0V
2.0%+2d
600V 2.5%+4d
NMRR: > 50dB @ 50Hz/60Hz
CMRR: > 120dB @ DC, 50Hz/60Hz; Rs=1k
Ω
Input Impedance: 10 M
Ω, 30pF nominal; (1000MΩ for
400.0mV range)
AC Voltage (AM-42)
RANGE Accuracy
50Hz -- 60Hz
4.000V, 40.00V, 400.0V
2.0%+5d
60Hz -- 500Hz
4.000V, 40.00V, 400.0V
3.0%+5d
50Hz -- 500Hz
600V 3.5%+5d
Input Impedance: 10 M
Ω, 30pF nominal
CMRR: > 60dB @ DC to 60Hz, Rs=1k
Ω
Capacitance (AM-45 & AM-47)
RANGE
1)
Accuracy
2)
100.0nF, 1000nF, 10.00
µF,
100.0
µF
3)
, 2000
µF
4)
3.5%+6d
5)
1)
Accuracy below 50nF is not specified
2)
Accuracies with film capacitor or better
3)
AM-45 top range. Updates > 1 minute on large
values
4)
AM-47 only. Updates > 1 minute on large values
5)
Specified with battery voltage above 2.8V (half full
battery). Accuracy decreases gradually to 12% at
low battery warning voltage of approx 2.4V
Capacitance (AM-42)
RANGE
1)
Accuracy
2)
500.0nF, 5.000
µF, 50.00µF,
500.0
µF, 3000µF
3)
3.5%+6d
4)
1)
Additional 50.00nF range accuracy is not specified.
2)
Accuracies with film capacitor or better.
3)
Updates > 1 minute on large values
4)
Specified with battery voltage above 2.8V (half full
battery). Accuracy decreases gradually to 12% at
low battery warning voltage of approx 2.4V
Ohms (AM-45 & AM-47)
RANGE
1)
Accuracy
600.0
Ω
2.0%+6d
2)
6.000k
Ω
1.2%+6d
2
)
60.00K
Ω, 600.0KΩ
1.0%+4d
6.000M
Ω
2.0%+4d
Open Circuit Voltage: 0.4VDC typical
1)
AutoCheck
TM
is for 6.000k
Ω ~ 6.000MΩ ranges;
600.0
Ω is an independent range for AM-47 only
2)
Add 40d to specified accuracy while reading is below
20% of range
Ohms (AM-42)
RANGE Accuracy
400.0
Ω
1.5%+6d
4.000K
Ω, 40.00KΩ, 400.0KΩ
1.0%+4d
4.000M
Ω
1.5%+4d
40.00M
Ω
2.5%+4d
Open Circuit Voltage: 0.4VDC typical
Hz Frequency (AM-45
1)
& AM-47
1) 2)
)
RANGE Accuracy
Specified
at
10.00Hz -- 30.00kHz
1) 2)
< 20V Sine-rms
10.00Hz -- 999.9Hz
2)
0.5%+4d
< 600V Sine-rms
Sensitivity (Sine-rms):
1)
Hz in Auto-V
Ω position: > 3V
2)
Line-level Hz in V position (AM-47 only):
@ 6.000VAC range: > 3V
@ 60.00VAC range: > 6V
@ 600.0VAC range: > 60V
Hz Frequency (AM-42)
RANGE
1)
Accuracy
5.000Hz -- 1.000MHz
0.5%+4d
1)
Specified at Input Voltage < 20VAC rms
Input Signal: Sine-wave, or Square-wave with duty
cycle > 40% & < 70%
Sensitivity (V-peak):
5Hz -- 100kHz
: > 1.3Vp
100kHz -- 500kHz : > 2.2Vp
500kHz -- 1MHz : > 4.2Vp
Diode Tester (AM-42 & AM-47 only)
Test Current (Typical)
Open Circuit Voltage
0.25mA for AM-42
0.48mA for AM-47
< 1.6VDC