Update Rate :
Power function: 1 per second nominal
Voltage, ACA clamp-on, Ohm, Hz & Temperature functions: 4 per second nominal
Polarity : Automatic
Low Battery : Below approx. 2.4V
Operating Temperature : 0
o
C to 40
o
C
Relative Humidity : Maximum relative humidity 80% for temperature up to 31
o
C
decreasing linearly to 50% relative humidity at 40
o
C
Altitude : Operating below 2000m
Storage Temperature : -20
o
C to 60
o
C, < 80% R.H. (with battery removed)
Temperature Coefficient : nominal 0.15 x (specified accuracy)/
o
C @(0
o
C -18
o
C or
28
o
C -40
o
C), or otherwise specified
Sensing : True RMS sensing for all models
Safety : Meets IEC61010-2-032 (1994), EN61010-2-032(1995), UL3111-2-032(1999).
Measurement Category : III 600 Volts ac & dc
Transient protection : 6.5kV (1.2/50
µs surge) for all models
Pollution degree : 2
E.M.C. : Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
an RF field of 3V/m:
otal Accuracy = Specified Accuracy + 45 digits
erformance above 3V/m is not specified
Overload Protections :
ACA Clamp-on jaws : AC 1000A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Power Supply : standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Power Consumption :
oltage, ACA, Hz & Power functions: 10mA typical
hm & Temperature functions: 4mA typical
APO Timing : Idle for 17 minutes
APO Consumption : 10
µA typical
Dimension : L224mm X W78mm X H40mm
Weight : 224 gm approx
Jaw opening & Conductor diameter : 45mm max
Special features : Backlight display (model ACD-31P & ACD-41PQ only); AutoVA
TM
(Auto Selection on ACV, DCV or ACA functions); Power measurement of selectable W,
VAR & VA with dual-display Total Power Factor features; Total harmonic distortion
THD%-F (model ACD-41PQ only); PEAK-rms HOLD