Calibration interval, Specifications – Amprobe AC50A User Manual
Page 12

Calibration Interval
We suggest a calibration interval of one year. If the instrument is used
very often or if it is used under rough conditions we recommend shorter
intervals. If the instrument is used only a few times a year, the calibration
interval can be extended to 3 years.
Specifications
(at 23° C % 5° C, max. 75 % rel. humidity)
Display:
3
3
/4
, LCD incl. functions and symbols
Bargraph:
40 segments
Range Selection:
manual
Auto-Power-Off:
approx. 30 min.
Overload display:
the left digit is blinking
Measuring rate:
20 measurements/sec. (Bargraph)
2 measurements/sec. (LCD)
Clamp opening:
ca. 30 mm (1/2 inch)
Overvoltage Categorie:
CAT II, 600 V
Pollution degree:
2
Height above MSL.:
up to 2000 m
Battery display:
at low battery
Power supply:
Batterie 2 x 1,5 V IEC LR6 (alkaline)
Current Consumtion:
approx. 10 mA
Operation temperature:
-10° C … 50° C
Storage temperature:
20° C … 60° C
Humidity:
< 75 % relative humidity
Dimension:
183 x 63,6 x 35,6 mm ( 7.2 x 2.5 x 1.4 inch)
Weight:
ca. 190 g (7 oz)
Current AC
Range
Resolution Accuray
50 Hz/60 Hz
40 Hz … 1 kHz
0 – 40 mA
10 µA
0 – 400 mA
100 µA
±(1,5 % rdg. +3 D) ±(2,0 % rdg. +5 D)
0 – 4 A
1 mA
0 – 40 A
10 mA
0 – 50 A
0,1 A
±(1,5 % rdg. +5 D) ±(2,0 % rdg. +5 D)
50 – 60 A
0,1 A
±(3,0 % rdg. +5 D) ±(3,5 % rdg. +5 D)
Voltage AC
Range
Resolution Accuray
50 Hz/60 Hz
40 Hz … 1 kHz
0...400 V
0,1 V
±(1,0 % rdg. +3 D) ±(2,0 % rdg..± 4 D)
10