beautypg.com

Elo TouchSystems AMM15TK2-TEW User Manual

Page 23

background image

23

This monitor is intended for use in Health Care Facilities model AMM15TK2

Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide.

No user serviceable parts inside, ask qualified personnel when accessing inside.

For disposal of waste product, follow the requirement of local code.

Electrical input rating: 12V DC 3.5A

Classification


Type of protection against electric shock: Class I Equipment.

Degree of protection against the ingress of water: IPX0

Mode of operation: Continuous

This monitor has been tested to comply with IEC/EC 60601-1 and IEC/EN60601-1-2 Certified
by UL to medical standard UL60601-1(UL/cUL Mark).

Because many medical offices are located in residential areas, this monitor, in addition to the
medical requirements, has also been tested and found to comply with the limits for FCC Class B
computing devices in a typically configured system. It is the system integrator or configurer’s
responsibility to test and ensure that the entire system complies with applicable EMC laws.
Environmental conditions for transport and storage:

-

Temperature range within –4

°

to 140

°

F (-20

°

to 60

°

C)

-

Relative humidity range within 10% to 90%

-

Atmospheric pressure range within 500 to 1060 hPa.

This manual is related to the following products: