Inlet filter, Returns, Disposal instructions – Precision Medical PM4351 EasyFlow5 User Manual

Page 11: Troubleshooting

background image

9

Inlet Filter:

1. Remove the Inlet Filter Cover by inserting the tip of a flat blade

screwdriver into the slot shown and prying off.

2. Grasp center of Inlet Filter Cap and pull to remove.

3. Remove Filter.

4. Wash Filter with mild detergent.

5. Rinse thoroughly with water and allow to thoroughly air dry.

6. Reinstall Filter back into the Inlet Filter Box.

7. Grasp center of Inlet Filter Cap and push into place.

8. Snap Inlet Filter Cover back into opening of the device cover.

Returns

Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns
must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods
damaged in transit. Refer to Precision medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.

Disposal Instructions

This device contains components that are hazardous to the environment. DO NOT dispose device into standard trash.
Contact your local waste management for disposal of Medical Electrical Devices.

Troubleshooting

Visual Indicator / Problem

Audible Alarm

Probable Cause

Remedy

YELLOW indicator flashes

Repeated
beeping

Flow to patient is occluded

Remove occlusion, kink or block in
patient cannula

When device is turned
“ON”, device does not
function and the GREEN
Power Indicator does not
light.

Continuous

Concentrator is not properly
plugged into electrical outlet.

Make sure Concentrator is properly
plugged into the electrical outlet.

RED General Alarm
Indicator is “ON” steady
or continuously blinking;
device stops functioning.

Repeated
beeping

System malfunction.

Turn power switch “OFF” and wait
5 minutes, then turn back “ON”. If
condition persists, turn the
concentrator “OFF”, connect to
another source of oxygen and
contact your Provider.