OPERATING INSTRUCTIONS

GENERAL

The function of the TIMOX Oxygen Timer is to

display and record the amount of time oxygen has

been flowing to a patient. The display is constantly

visible, indicates when oxygen is flowing, and is

easily reset.

OPERATION

Place the TIMOX Oxygen Timer in service

by connecting its handtight inlet fitting to the

downstream side of a flowmeter or regulator. The

timer generates a small amount of backpressure;

therefore it should be used with a backpressure

compensated flowmeter. Besides oxygen, the

timer can be used with compressed air, helium,

nitrogen, or carbon dioxide.

DO NOT USE IN HAZARDOUS ATMOSPHERES

.

READING HOURS

The display is an LCD and is constantly visible.

The timer will start running when the flow reaches

.4 LPM ±.1 LPM. The decimal point blinks when

the timer is running. The display is capable of

showing a maximum elapsed time of 1999.9 hours.

RESETTING

To reset the timer to 0 hours, simply insert a

pen or pencil tip into the hole marked “RESET” on

the front of the timer to press the reset switch.

CARE

The TIMOX Oxygen Timer is an electronic

device and must be handled with reasonable care.

Do not submerge the unit in any liquid. Do not au-

toclave it. Clean the exterior with a mild detergent

and a damp cloth.