beautypg.com

O2 sensor, Functionality, Physical location – Ion Science SF6 P2 User Manual

Page 12: Sensor type, Sensor lifetime, User interface, Safety

background image

SF6 P1 & P1:p MANUAL

Ion Science Ltd

11 of 15

Unrivalled Detection. www.ionscience.com

O2 sensor


Functionality
The optionally available O2 sensor continuously measures the O2 content in the ambient air, and reports it to
the network controller if present.

An alarm condition on both the measuring station and network controller is raised whenever the oxygen level
is outside the allowed range of 16…25%.

Physical location
The O2 sensor is part of the SmartSensor, it is found protruding downwards from it.

Sensor Type
The O2 sensor is of galvanic cell type. It exhibits low cross sensitivity to CO2 and other gases, good stability
and linearity over its range.

Sensor Lifetime
With a reasonable safety margin, sensor life is fixed at one year, no matter if shelfed or used

– this is due to

the nature of galvanic cells, in which the chemical reaction goes on at all times as long as oxygen is present.
When lifetime is over, the SmartSensor must be replaced against a fresh, readily calibrated one.

User Interface
When a low oxygen condition is encountered, the measuring station will issue an alarm, involving the built in
alarm buzzer, the red alarm lamp, and the alarm relay. The low oxygen condition will be reflected by the
yellow "500" and "1000" ppm lamps flashing.
In case of a high oxygen condition, the same applies, except that the "1500" and "2000" ppm lamps
will flash.
On the SF6 NETWORK CONTROLLER P2NC, the O2 status will be marked as "OK", "HI", or "LO",
respectively, for each measuring station.

Safety
Should the O2 sensor fail, be missing or spent, a low oxygen alarm will be issued.