5

Installation

Remove the grub screws on the drive front panel to enable the pumphead to be fitted.

Fit the track in any one of three orientations, over the drive shaft and locating boss. Secure the track with the locating screw.

Ensure the drive shaft is degreased before locating the rotor onto the shaft via the split collet. Tighten the rotor screw to a torque

of 3Nm to prevent the collet slipping during operation.
To reposition the track, swing out the crank handle to expose the rotor retaining screw. Turn the screw anticlockwise one turn to

release the collet, and withdraw the rotor from the shaft. Loosen the track locating screw, and pull the track clear. Rotate the

track to its new position and tighten the track locating screw. Use this method of removal and fitting if cleaning is required.

Tube loading

Isolate pump from mains supply. Unlock and open the hinged guard and swing out the rotor crank handle until it locks into

position. Select the length of tubing required, noting that approximately 240mm is required for the track systems.
Fit one end of the tubing into one of the spring loaded clamps, and then, whilst rotating the rotor with the crank handle, feed the

tubing between the rollers and the track, aligning it within the rotor tube guides. The tubing must lie naturally against the track

and must not be twisted or stretched.

Fit the other end of the tubing into the second spring loaded clamp, ensuring that the tubing is not slack in the pumphead, since

this can reduce tube life.
Close the crank handle and shut and lock the guard.
After the pump has been started, open the delivery clamp for a short time, so that the tube can find its natural length.
The 501RL pumphead is fitted with four-position tube clamps, to accommodate various tube diameters, which can be adjusted

by pushing in or pulling out the bars at the top of the upper clamp and the bottom of the lower clamp. Set the clamps so that the

minimum necessary pressure is applied to the tubing.