7

Speed
A remote potentiometer with a nominal value of between 4.7kohm and 5 kohm should be wired as shown, When
using a remote potentiometer do not connect a voltage/current control signal at the same time. The speed control
signal will require calibration relative to the minimum and maximum settings of the potentiometer. Use the offset
and range potentiometers as described under Calibration for auto control.

Tachometer output
This facility can be used to indicate motor speed or total the number of motor revolutions. Select either 1) 0-5V DC
or 2) 5V pulse train output using the tachometer output switch.

Car

Car

Car

Car

Care and maintenance

e and maintenance

e and maintenance

e and maintenance

e and maintenance

The only scheduled maintenance of the pump unit is to inspect the motor brushes and to replace them before their
length is less than 10mm. The life of the brushes depends on the duty of the pump but is expected to be at least
4,000 hours at maximum speed.

If the pump requires cleaning use a mild solution of detergent in water after removing the pumphead. Do not use
strong solvents.

For gearbox rebuilds, use 15ml of the recommended lubricant RD-105. This is a SAE 30 mineral oil loaded with
molybdenum disulphide to form a soft fluid grease.

Specif

Specif

Specif

Specif

Specifica

ica

ica

ica

ication

tion

tion

tion

tion

Maximum rotor speed

55rpm, 220rpm, 300rpm

Shaft torque

2.2Nm

Control range

100:1

Voltage/frequency

100-120/220-240V 50/60Hz

Power consumption

100VA

Operating temperature range

5C to 40C

Storage temperature range

-40C to 70C

Noise

<70dB(A) at 1m

Weight

9.0kg

Standards

EN60529 (IP55)

Machinery Directive 98/37/EC EN60204-1

Low Voltage Directive 73/23/EEC EN61010-1

EMC Directive 89/336/EEC EN50081-1/EN50082-1