Fit the track over the gearbox boss.

Align the track horizontally so that the location holes are aligned with the threaded

gearbox holes.

Tighten the four track retaining screws using a Number 2 Pozi-Driv screwdriver.

Re-connect the guard interlock controlled waste pipework if required.

620RE, 620RE4 and 620R CIP and SIP

General

Unlock the guard and disengage the rollers within the tube zone.

Close the guard and squeeze against the track until the latch clicks.

Observe a 1m safety area.

CIP

LoadSure tube elements and continuous tubing can be cleaned using CIP processes.

Ensure that the tubing material is chemically compatible with the cleaning agent that

is to be used.

If cleaning agents are spilled over the pumphead, wash down immediately.

Ensure that controlled waste pipework is fitted to allow a safe release of cleaning

agent in the event of a tube failure.

SIP

Only STA-PURE tube elements can be used in a steam in place sterilisation

processes.

STA-PURE tubing elements can be sterilised to 3A Class two and FDA minimum

recommended standard which is 121C (250F) at 1bar (14.5 psi) saturated steam for

20 minutes.

Monitor the process continuously .

If a tube failure occurs, shut down the process. Do not touch the pumphead until a

20 minute cooling period has been observed.

Ensure a 20 minute acclimatisation period is observed before running the pump

following SIP.

Ensure that controlled waste pipework is fitted to allow a safe release of steam in

the event of a tube failure.

Ensure a 1m safety zone is maintained around the pumphead during SIP cycles.