Coil excitation electrodes, Pair 1 pair 2 – Tecfluid Cables For FLOMAT Series User Manual

Page 5

background image

The Pair 1 cables are connected to terminals 1, 2 & 3 for the coil excitation. The Pair 2 cables, for
the electrode inputs, are connected to terminals 17, 18, 19 & 20. Terminals 18 & 19 must be short
circuited together.

To connect to the sensor, first loosen the cable gland to allow the cable to slide in it.

Make sure that the two fixing screws of the cover do not protrude inside the cover (this avoids
damaging the O ring of the sensor neck).

Apply a little Vaseline on the sensor neck to aid inserting in the cover, specially on the O
rings.

Connect the two cable connectors in their corresponding male connectors on the sensor,
mating the bumps in the guide of the male connectors.

Slide the cover on the sensor neck until it meets its stop.

Tighten the two fixing screws to anchor the cover.

Tighten the cable gland to guarantee water tightness.

Connection to a Flomid MX converter.

5

Coil Excitation

Electrodes

Red (1)

Black (2)

Screen (3)

White (17)

Screen (18)

Shortcircuit (18 to 19)

Black (20)

Pair 1

Pair 2