Pouring gradients from the top
Filling a container with a gradient (e.g. casting
acrylamide pore gradient gels) can be done either
dense solution first (“from the top”), or light
solution first (“from the bottom”).
To fill from the top:
1
Proceed as described in Section 1, putting light (final
or top) solution in the reservoir chamber (step 3) and
dense solution in the mixing chamber (step 5).
2
Place the delivery outlet against the upper edge of
the receiving container. Adjust the pump rate so that
the solution flows evenly down the side in a smooth,
continuous stream. The delivery speed should be slow
enough that the newly arriving solution does not mix
with the underlying solution.
Alternatively, using a rigid cannula at the end of the
delivery tubing, hold the tip of the cannula just above
the surface of the solution, raising it smoothly as the
container fills.
- IEF100 (65 pages)
- IEF100 (65 pages)
- SE900 (37 pages)
- SE260 (31 pages)
- SE260 (31 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE410 (51 pages)
- SE410 (50 pages)
- SE640 (51 pages)
- SE640 (50 pages)
- SE275 (21 pages)
- SE245 (13 pages)
- SE675 (27 pages)
- SG100 (9 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- HE33 (26 pages)
- HE33 (26 pages)
- HE99X (26 pages)
- HE99X (26 pages)
- HE-PLUS System (26 pages)
- HE-PLUS System (26 pages)
- SUB Series (31 pages)
- SUB Series (31 pages)
- TE70X (31 pages)
- TE70X (31 pages)
- PR648 (11 pages)
- PR150 (9 pages)
- PR645 Immunoblot XL (21 pages)
- TE22 (35 pages)
- TE22 (35 pages)
- TE62 (35 pages)
- TE22 (31 pages)
- DQ300 (30 pages)
- DQ300 (30 pages)
- UVIS20 (21 pages)
- UVIS20 (21 pages)
- SP-2001 Vision Life Spectrophotometer (66 pages)
- Single Channel Variable Pipette (8 pages)
- PR250 (18 pages)
- PP24 (10 pages)